the use of pacemakers in infants and children

Clicks: 131
ID: 172759
2005
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #162 of 231 articles by views in planta medica

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 231 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract

Background: To review retrospectively our experience with pediatric cardiac pacing with an emphasis on specific problems in pacemaker implantation in infants and children.

Methods: Between April 1977 and March 2005 ninety-one child underwent permanent pacemaker implantation, 46 boys (50.5%) and 45 girls (49.5) with a mean age of 7.9 years (the oldest 17 years and the youngest 1 day). The age of the children at the time of implantation was 2 to 9 years in one half and 14 to 17 years in one fourth of them. Five children were younger than one month and three of them younger than one week. The indications for permanent pacing were complete atrioventricular block (CAVB) in more than half patients, sick sinus syndrome in one fourth and CAVB following open heart surgery in nine. The mean time for elective battery replacement was 6.5 years after the implantation. In one third of the children the battery was replaced before the end of the fifth year after the implantation and after seven years in more than half of the patients. Epicardial leads and left anterior thoracotomy or subxyphoid approach were used at the beginning. From the year 1980 we prefered transvenous leads. Modern epicardial steroideluting leads demonstrated relatively stable acute and chronic pacing and sensing tresholds. VVI/R pacemaker was the pacemaker of choice at the first implantation and we upgraded the pacing at the time of the elective battery replacement. Children with VDD pacemakers needed sometimes upgrading of pacing sooner because of the inappropriate atrial sensing and AV synchronyzation.

Conclusions: Better quality pacing leads and smaller but more powerful pacemaker generators with autocapturecontrolled pacing options became more and more attractive for our little patients because of better implantation facilities, substantial energy savings and battery life prolongation.

Reference Key
kosin2005zdravnikithe Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Miro Kosin
Journal planta medica
Year 2005
DOI
DOI not found
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.