h2s interference on co2 isotopic measurements using a picarro g1101-i cavity ring-down spectrometer

Clicks: 173
ID: 161163
2015
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #91 of 183 articles by views in bioorganic & medicinal chemistry

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 183 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Cavity ring-down spectrometers (CRDSs) have the capacity to make isotopic measurements of CO2 where concentrations range from atmospheric (~ 400 ppm) to 6000 ppm. Following field trials, it has come to light that the spectrographic lines used for CO2 have an interference with elevated (higher than ambient) amounts of hydrogen sulfide (H2S), which causes significant depletions in the δ13C measurement by the CRDSs. In order to deploy this instrument in environments with elevated H2S concentrations (i.e., active volcanoes), we require a robust method for eliminating this interference. Controlled experiments using a Picarro G1101-i optical spectrometer were done to characterize the H2S interference at varying CO2 and H2S concentrations. The addition of H2S to a CO2 standard gas reveals an increase in the 12CO2 concentration and a more significant decrease in the 13CO2 concentration, resulting in a depleted δ13C value. Reacting gas samples containing H2S with copper prior to analysis can eliminate this effect. Models post-dating the G1101-i carbon isotope analyzer have maintained the same spectral lines for CO2 and are likely to have a similar H2S response at elevated H2S concentrations. It is important for future work with CRDS, particularly in volcanic regions where H2S is abundant, to be aware of the H2S interference on the CO2 spectroscopic lines and to remove all H2S prior to analysis. We suggest employing a scrub composed of copper to remove H2S from all gas samples that have concentrations in excess of 1 ppb.
Reference Key
malowany2015atmospherich<sub>2</sub>s Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;K. Malowany;J. Stix;A. Van Pelt;G. Lucic
Journal bioorganic & medicinal chemistry
Year 2015
DOI
10.5194/amt-8-4075-2015