Cadaveric biomechanical testing of torque - to - failure magnitude of Bilateral Apical Vertebral Derotation maneuver in the thoracic spine.

Clicks: 277
ID: 15158
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #1,098 of 2,056 articles by views in PloS one

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 2,056 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
It remains unclear what is the real safe limit of torque magnitude during Bilateral Apical Vertebral Derotation (BAVD) in thoracic curve correction. Up to author's knowledge there is no study except this one, to reproduce in-vivo real measurements and intraoperative conditions during BAVD maneuver. The objective of this study was to evaluate the torsional strength of the instrumented thoracic spine under axial rotation moment as well as to define safety limits under BAVD corrective maneuver in scoliosis surgery. 10 fresh, full-length, young and intact human cadavers were tested. After proper assembly of the apparatus, the torque was applied through its apical part, simulating thoracic curve derotation. During each experiment the torque magnitude and angular range of derotation were evaluated. For more accurate analysis after every experiment the examined section of the spine was resected from the cadaver and evaluated morphologically and with a CT scan. The average torque to failure during BAVD simulation was 73,3 ± 5,49Nm. The average angle of BAVD to failure was 44,5 ± 8,16°. The majority of failures were in apical area. There was no significant difference between the fracture occurrence of left or right side of lateral wall of the pedicle. There was no spinal canal breach and/or medial wall failure in any specimen. The safety limits of thoracic spine and efficacy of BAVD for axial plane correction in the treatment of Adolescent Idiopathic Scoliosis (AIS) were established. It provided qualitative and quantitative information essential for the spinal derotation under safe loading limits.
Reference Key
pankowski2019cadavericplos Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Pankowski, Rafal;Roclawski, Marek;Ceynowa, Marcin;Mazurek, Tomasz;Ciupik, Lechoslaw;Kierzkowska, Agnieszka;
Journal PloS one
Year 2019
DOI
10.1371/journal.pone.0221494
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.