comprehensive laboratory and field testing of cavity ring-down spectroscopy analyzers measuring h2o, co2, ch4 and co
Clicks: 207
ID: 143658
2015
Article Quality & Performance Metrics
Overall Quality
Improving Quality
0.0
/100
Combines engagement data with AI-assessed academic quality
Reader Engagement
Popular Article
30.0
/100
205 views
38 readers
Trending
AI Quality Assessment
Not analyzed
Abstract
To develop an accurate measurement network of greenhouse gases,
instruments in the field need to be stable and precise and thus
require infrequent calibrations and a low consumption of
consumables. For about 10 years, cavity ring-down spectroscopy
(CRDS) analyzers have been available that meet these stringent
requirements for precision and stability. Here, we present the
results of tests of CRDS instruments in the laboratory (47
instruments) and in the field (15 instruments). The precision and
stability of the measurements are studied. We demonstrate that,
thanks to rigorous testing, newer models generally perform better
than older models, especially in terms of reproducibility between
instruments. In the field, we see the importance of individual
diagnostics during the installation phase, and we show the value of
calibration and target gases that assess the quality of the
data. Finally, we formulate recommendations for use of these
analyzers in the field.
| Reference Key |
kwok2015atmosphericcomprehensive
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;C. Yver Kwok;O. Laurent;A. Guemri;C. Philippon;B. Wastine;C. W. Rella;C. Vuillemin;F. Truong;M. Delmotte;V. Kazan;M. Darding;B. Lebègue;C. Kaiser;I. Xueref-Rémy;M. Ramonet |
| Journal | bioorganic & medicinal chemistry |
| Year | 2015 |
| DOI |
10.5194/amt-8-3867-2015
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.