comparison of two vhit software programs in diagnostics of semicircular canals
Clicks: 209
ID: 140558
2017
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
209 views
35 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #2 of 2 articles by views in texto & contexto enfermagem
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Introduction: The aim of this work was to assess SCC function in VHIT test, using both versions of VHIT ULMER.
Material and methods: The study was performed in 57 subjects aged 22–33 years (48 women and 9 men) without vertigo in anamnesis. Patients did not complain of any “vertigo” ailments or any balance disorder incidents in the past. All tests using either VHIT I or VHIT II were performed by the same person. The function of lateral SCC in VHIT ULMER II was examined similarly like in VHIT ULMER I. In both methods was used the same sensitivity threshold parameters for horizontal and vertical accelerations (horizontal – 2500, vertical – 1600). Each of these sequences enables a parameter, called represented GAIN, to be calculated in form of a point on the canalogram.
Results : GAIN represents the deviation of the gaze in comparison to the deviation of the head between the first (t – 0 ms) and the fourth image (t = 120 ms). The calculation formula is: GAIN (in %) =100 x DG/RH. DG is deviation of the gaze (“gaze velocity in space”) and RH is rotation of the head (“head velocity or impulse canal paresis”). Normal GAIN value ranges 1–40%.
Conclusions : The examination using VHIT ULMER II appears to be a more sensitive diagnostic method than VHIT ULMER I. Improved software with automated functions such as camera adjustments, examination surface arrangement, of VHIT ULMER II, enables technician to shorten the time of examination, simultaneously providing a range of new information about the condition of oculo-vestibular system.
| Reference Key |
pietkiewicz2017archivescomparison
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Piotr Pietkiewicz;Jaroslaw Milonski;Marzena Bielinska;Jurek Olszewski |
| Journal | texto & contexto enfermagem |
| Year | 2017 |
| DOI |
10.5114/aoms.2017.66048
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.