the influence of temperature calibration on the oc–ec results from a dual-optics thermal carbon analyzer
Clicks: 254
ID: 140370
2014
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
254 views
37 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #28 of 183 articles by views in bioorganic & medicinal chemistry
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 183 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Thermal–optical analysis (TOA) is a widely used technique that fractionates
carbonaceous aerosol particles into organic and elemental carbon (OC and EC),
or carbonate. Thermal sub-fractions of evolved OC and EC are also used for
source identification and apportionment; thus, oven temperature accuracy
during TOA analysis is essential. Evidence now indicates that the "actual"
sample (filter) temperature and the temperature measured by the built-in oven
thermocouple (or set-point temperature) can differ by as much as
50 °C. This difference can affect the OC–EC split point selection and
consequently the OC and EC fraction and sub-fraction concentrations being
reported, depending on the sample composition and in-use TOA method and
instrument. The present study systematically investigates the influence of an
oven temperature calibration procedure for TOA. A dual-optical carbon
analyzer that simultaneously measures transmission and reflectance (TOT and
TOR) is used, functioning under the conditions of both the National Institute of Occupational Safety and
Health Method 5040 (NIOSH) and
Interagency Monitoring of Protected Visual Environment (IMPROVE) protocols. The application of the oven calibration procedure to our dual-optics instrument significantly changed NIOSH 5040 carbon fractions (OC and
EC) and the IMPROVE OC fraction. In addition, the well-known OC–EC split
difference between NIOSH and IMPROVE methods is even further perturbed
following the instrument calibration. Further study is needed to determine if
the widespread application of this oven temperature calibration procedure
will indeed improve accuracy and our ability to compare among carbonaceous
aerosol studies that use TOA.
| Reference Key |
pavlovic2014atmosphericthe
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;J. Pavlovic;J. S. Kinsey;M. D. Hays |
| Journal | bioorganic & medicinal chemistry |
| Year | 2014 |
| DOI |
10.5194/amt-7-2829-2014
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.