[Design and validation of calibration between tube focus spot and center plane of rotation in computed tomography system].

Clicks: 281
ID: 13727
2019
Article Quality & Performance Metrics
Overall Quality Improving Quality
0.0 /100
Combines engagement data with AI-assessed academic quality
AI Quality Assessment
Not analyzed
Abstract
This study proposed a method to calibrate tube focus spot and the center plane of rotation in computed tomography system. In the method, the tube was rotated to 0° and 180° respectively, and then one metal jig with symmetric windows A and B was scanned at each position under the tube cool and static condition. According to the geometry of tube focus spot, aperture center of the collimator and jig, the distance between tube focus spot and the center plane of rotation were calculated with the X ray transmittance data after denoising, mean value and normalization. To verify the practicability and validity of the method, the tube focus spot in a 16 slices CT system (Brivo CT385, GE, China) was calibrated, and the result after calibration was validated by scanning a polaroid film. The validation result showed that the deviation between tube focal spot and center plane of rotation was 0.02 mm and was in the error range within ± 0.1 mm. The results of this study showed that, as a simple and low-cost design, the method could be used for fast calibration between tube focus spot and the center plane of rotation.
Reference Key
yu2019designsheng Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Yu, Xiaomin;Yin, Wei;Lan, Yu;Liu, Zhihong;
Journal sheng wu yi xue gong cheng xue za zhi = journal of biomedical engineering = shengwu yixue gongchengxue zazhi
Year 2019
DOI
10.7507/1001-5515.201812016
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.