comparing the corneal curvatures obtained from three different keratometers-iol master, bausch & lomb manual keratometer and topcon kr-8800 autokeratometer
Article Quality & Performance Metrics
Readership in this journal
EmergingRanked #120 of 133 articles by views in journal of fungi
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 133 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
Abstract
METHODS: In this prospective study, keratometry was performed using an IOL Master(IM), a Bausch & Lomb manual keratometer(Man)and TOPCON KR-8800 autokeratometer(Top)on 252 eyes of patients recruited from camps for cataract surgery. The average keratometry values were recorded and compared. The agreements between the instruments were analyzed using the Bland-Altman statistical method. The main outcome measure was average keratometry values.
RESULTS:1)IOL Master and Manual keratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.60±1.52 D with the Manual keratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and Manual keratometer(P=0.001). The 95% LOAs of the two devices were -0.22 to 0.22 as shown in the Bland-Altman plot; 2)IOL Master and autokeratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.46±1.53 D with the autokeratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and autokeratometer(P<0.0001). The 95% LOAs of the two devices were -0.24 to 0.55 as shown in the Bland-Altman plot; 3)Autokeratometer and Manual keratometer: the mean corneal power was 44.60±1.52 D with Manual keratometer and 44.46±1.53 D with the autokeratometer. The paired ttest demonstrated a statistically significant difference in the mean corneal power between the autokeratometer and Manual keratometer(P<0.0001). The 95% LOAs of the two devices were -0.30 to 0.57 as shown in the Bland-Altman plot.
CONCLUSION: Keratometry data obtained with different instruments may not be interchangeable, a fact that has important implications for cataract surgeons with respect to both surgical planning and outcomes auditing.
| Reference Key |
pereira2018guojicomparing
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Arpitha Pereira;Ajita Sasidharan |
| Journal | journal of fungi |
| Year | 2018 |
| DOI |
10.3980/j.issn.1672-5123.2018.1.04
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.