comparing the corneal curvatures obtained from three different keratometers-iol master, bausch & lomb manual keratometer and topcon kr-8800 autokeratometer

Clicks: 172
ID: 135914
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #120 of 133 articles by views in journal of fungi

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 133 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
AIM: To compare the corneal curvature and to investigate the agreement between three different keratometers.

METHODS: In this prospective study, keratometry was performed using an IOL Master(IM), a Bausch & Lomb manual keratometer(Man)and TOPCON KR-8800 autokeratometer(Top)on 252 eyes of patients recruited from camps for cataract surgery. The average keratometry values were recorded and compared. The agreements between the instruments were analyzed using the Bland-Altman statistical method. The main outcome measure was average keratometry values.

RESULTS:1)IOL Master and Manual keratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.60±1.52 D with the Manual keratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and Manual keratometer(P=0.001). The 95% LOAs of the two devices were -0.22 to 0.22 as shown in the Bland-Altman plot; 2)IOL Master and autokeratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.46±1.53 D with the autokeratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and autokeratometer(P<0.0001). The 95% LOAs of the two devices were -0.24 to 0.55 as shown in the Bland-Altman plot; 3)Autokeratometer and Manual keratometer: the mean corneal power was 44.60±1.52 D with Manual keratometer and 44.46±1.53 D with the autokeratometer. The paired ttest demonstrated a statistically significant difference in the mean corneal power between the autokeratometer and Manual keratometer(P<0.0001). The 95% LOAs of the two devices were -0.30 to 0.57 as shown in the Bland-Altman plot.

CONCLUSION: Keratometry data obtained with different instruments may not be interchangeable, a fact that has important implications for cataract surgeons with respect to both surgical planning and outcomes auditing.

Reference Key
pereira2018guojicomparing Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Arpitha Pereira;Ajita Sasidharan
Journal journal of fungi
Year 2018
DOI
10.3980/j.issn.1672-5123.2018.1.04