Microextraction in packed syringe (MEPS) for liquid and gas chromatographic applications. Part II--Determination of ropivacaine and its metabolites in human plasma samples using MEPS with liquid chromatography/tandem mass spectrometry.

Clicks: 271
ID: 104897
2004
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #2 of 3 articles by views in journal of mass spectrometry : jms

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
A new technique for sample preparation on-line with liquid chromatographic/tandem mass spectrometric (LC/MS/MS) assay was developed. Microextraction in a packed syringe (MEPS) is a new miniaturized, solid-phase extraction technique that can be connected on-line to gas or liquid chromatography without any modifications. In MEPS approximately 1 mg of the solid packing material is inserted into a syringe (100-250 microl) as a plug. Sample preparation takes place on the packed bed. The bed can be coated to provide selective and suitable sampling conditions. The new method is very promising, very easy to use, fully automated, of low cost and rapid in comparison with previously used methods. This paper presents the development and validation of a method for MEPS on-line with LC/MS/MS. Ropivacaine and its metabolites (PPX and 3-OH-ropivacaine) in human plasma samples were used as model substances. The method was validated and the calibration curves were evaluated by means of quadratic regression and weighted by the inverse of the concentration, 1/x, for the calibration range 2-2000 nM. The applied polymer could be used more than 100 times before the syringe was discarded. The extraction recovery was between 40 and 60%. The results showed high correlation coefficients (R(2) > 0.999) for all analytes in the calibration range studied. The accuracy, expressed as a percentage variation from the nominal concentration values, ranged from 0 to 6%. The precision, expressed as the relative standard deviation, at three different concentrations (quality control samples) was consistently about 2-10%. The limit of quantification was 2 nM.
Reference Key
abdelrehim2004microextractionjournal Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Abdel-Rehim, Mohamed;Altun, Zeki;Blomberg, Lars;
Journal journal of mass spectrometry : jms
Year 2004
DOI
DOI not found
URL URL not found
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.