The paper describes the design and implementation of an infrared (IR) thermometer with automatic emissivity correction. The temperature measurement is carried out by the simple digital thermopile sensor MLX90614. The emissivity correction is based on benefits of diffuse reflecting materials and it uses an IR laser diode in conjunction with a selective amplifier. Moreover, the paper includes the design of the control interface with a graphics LCD. Furthermore, this paper describes the power supply unit with a Li-ion cell controlled by basic integrated circuits.Reference Key |