Determination of oxygen r at 3 Tesla using samples with a concentration range of dissolved oxygen.

Clicks: 199
ID: 60940
2019
Article Quality & Performance Metrics
Overall Quality Improving Quality
0.0 /100
Combines engagement data with AI-assessed academic quality
AI Quality Assessment
Not analyzed
To investigate the sensitivity of modified Look-Locker inversion recovery (MOLLI) to measure changes in dissolved oxygen (DO) concentrations in water samples and to calculate sequence-specific relaxivity (r) and limit of detection (LOD).Ten water samples with a range of DO concentrations were scanned at 3 T using two variations of MOLLI schemes. Using linear regression the r of DO was estimated from the measured DO concentrations and T relaxation rates (R). The results were combined with previously reported values on in vivo stability measures of the MOLLI sequences and used to estimate a LOD.DO concentrations ranged from 0.5 to 21.6 mg L. A linear correlation between DO and R was obtained with both MOLLI sequences, with an average correlation coefficient (R) 0.9 and an average estimated r ([Formula: see text]) of 4.45 × 10 s mg L. Estimated LOD was ≈ 10 mg L.MOLLI T-mapping sequences may be used for detecting dissolved oxygen in vivo at 3 T with an [Formula: see text] in the range 4.18-4.8 × 10 s mg L and a corresponding LOD for dissolved oxygen of approximately 10 mg L. MOLLI-based T mapping may be a useful non-invasive tool for quantification of in vivo changes of DO concentration during oxygen challenges.
Reference Key
vatnehol2019determinationmagma Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Vatnehol, Svein Are Sirirud;Hol, Per Kristian;Bjørnerud, Atle;Amiry-Moghaddam, Mahmood;Haglerød, Camilla;Storås, Tryggve Holck;
Journal magma (new york, ny)
Year 2019
DOI 10.1007/s10334-019-00783-x
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.